Abstract
This application note describes the use of the Kairos Amino Acid Kit for analysis of20 amino acids in solution.Kairos Amino Acid Kit enables biomedical researchers to achieve trusted results through a single flexible kit that allows them to accurately quantitate 20 amino acids within their normal physiological range in less than10 minutes using the ACQUITY UPLC/ACQUITY QDa System.
Benefits
- Analytical selectivity of the chromatographic method provides separation of isobaric species
- Fast analytical run times (<10 mins)
- Confidence in peak detection
- Greater flexibility, system can be used for other analyses
Introduction
The research only Kairos Amino Acid Kit (p/n: 176004379) is designed for biomedical researchers to enable the analysis of up to 45 biologically relevant amino acids in less than 10 minutes. The kit calibrators have been value assigned using higher order reference material where available to provide added confidence in your results. Metrological traceability of the Kairos Amino Acid Kit Calibrators (p/n: 186009048) were traceable to National Metrology Institute of Japan (NMIJ CRM 6011a-6018a, 6022a) for 9 amino acids and a further 8 were traceable to National Instituteof Standards and Technology (NIST SRM 2389a), the remaining amino acids were gravimetrically prepared using TraceCERT standards from Sigma-Aldrich.
Here we describe the use of the Kairos Amino Acid Kit for analysis of20 amino acids in solution. Chromatographic separation was performed using an ACQUITY UPLC H-Class System using a CORTECS UPLC C18 1.6 µm, 2.1 × 150 mm Column (p/n: 186007096), followed by detection on an ACQUITY QDa Mass Detector, designed for simplicity with automated set-up (Figure 1). Performance of the kit and analytical system was assessed in solution and NIST SRM 2389a material analyzed for biomedical research purposes.
The chromatographic conditions used allowed for the separation of leucine and alanine isobaric species.
Figure 1. The Waters ACQUITY UPLC H-Class System/ACQUITY QDa Mass Detector System.
Experimental
Reagent Kit
Kairos Amino Acid Kit (p/n: 176004379)
LC conditions
System: | ACQUITY UPLC H-Class Column Heater (CH-A) |
Needle: | 15 μL |
Column: | CORTECS UPLC C18 2.1 x 150 mm, 1.6 μm (p/n: 186007096) |
Pre-column: | ACQUITY UPLC Column In-line Filter Kit (p/n: 205000343) |
Mobile phase A: | Water with 0.1% formic acid |
Mobile phase B: | Acetonitrile with 0.1% formic acid |
Needle wash solvent: | Mobile phase B |
Purge solvent: | Mobile phase A |
Column temp.: | 55 °C |
Injection volume: | 2 μL |
Flow rate: | 0.5 mL/min |
Gradient: | See Table 1 |
Run time: | 9 minutes |
MS conditions
System: | ACQUITY QDa Mass Detector |
Acquisition mode: | Single Ion Recording (SIR) (see Tables 2 and 3 for details) |
Polarity: | ESI+ |
Capillary: | 0.8 kV |
Source temp.: | 120 °C |
Desolvation temp.: | 600 °C |
Cone: | 10 V |
Data management
MassLynx v4.1 Software with TargetLynx Application Manager
Kit reconstitution
To reconstitute the freeze-dried Kairos Amino Acid Kit, follow the Care and Use Manual (720005448EN). Calibrators and QCs, were reconstituted using 2 mL of 0.1 M HCl and mixed at room temperature for a minimum of 30 minutes, ensuring all material is fully dissolved.
Internal Standard was reconstituted using 2 mL of water and mix at room temperature for 10 minutes, ensuring all material is fully dissolved. The contents of the vial were transferred to a volumetric flask and make up to 10 mL using 10% sulfosalicylic acid supplied in the kit.
Reagent preparation
To prepare the Kairos Amino Acid Kit reagents, Borate buffer and AccQ•Tag Ultra “3X” Derivitization Kit reagent, follow the Care and Use Manual (720005448EN). Borate buffer preparation: Add 430 µL of 0.5 M NaOH aq. to the 6 mL of Borate buffer provided in the derivatization kit (Reagent 1) Expiry: 3 months. Storage: Room temperature.
AccQ•Tag Ultra “3X” reagent: Reconstitute AccQ•Tag Ultra“3X” (Reagent 2B) in 1.5 mL acetonitrile (Reagent 2A). Heat for 10 minutes at 55 °C and vortex. Expiry: 5 days. Storage: Room temperature. AccQ•Tag Ultra “3X” should be stored in the desiccator once reconstituted.
Sample preparation
Step 1 | Add 50 μL of sample to 1.5 mL Eppendorf |
Step 2 | Add 50 μL of Internal Standard |
Step 3 | Vortex mix for 5 seconds |
Step 4 | Add 50 μL of water |
Step 5 | Vortex mix for 5 seconds |
Step 6 | For high concentration samples only – Add 1000 μL of diluent (0.1 M HCl) |
Step 7 | Centrifuge for 15 minutes at 9000 g |
Step 8 | Add 70 μL of Borate buffer to maximum recovery vial |
Step 9 | Add 10 μL of supernatant into Borate buffer and pipette mix |
Step 10 | Add 20 μL of AccQ-Tag Ultra “3X” reagent |
Step 11 | Vortex for 5 seconds |
Step 12 | Allow sample to stand at room temperature for 1 min |
Step 13 | Heat for 10 minutes at 55 °C |
Step 14 | Inject 2 μL |
For precision and accuracy studies, panel samples were gravimetrically prepared in 0.1 M HCl at 20, 150, 400,700 µM for all amino acids and an additional high concentration panel at 2500 µM was prepared for alanine, glycine, isoleucine, leucine, phenylalanine, valine, serine, threonine, tyrosine, and glutamine.
Method conditions
Table 1. Gradient table for the separation of the amino acids.
Table 2. SIR parameters for the amino acids and calibrator concentration range.
Table 3. SIR parameters for the internal standards.
Results and Discussion
No significant system carryover (detector response was ≤20% of Calibrator 1) was observed from high concentration samples into subsequent blank injections.
Precision was determined by preparing and analyzing the panel samples in replicates of five over three separate days (n=15), within day and between day precision performance (%CV) were calculated for each panel. Each panel within day precision was ≤12.0%CV and the between day precision performance was ≤14.7%CV and the mean within day and between day precision performance was less than ≤11.3% CV as shown in Table 4.
Table 4. Mean within-day and between day precision for the analysis of 20 amino acids in solution.
To assess linearity, the Kairos Amino Acid Kit Calibrators were analyzed each day over three days. Regression analysis demonstrateda linear fit using 1/x weighting across the concentration range for the 20 amino acids analyzed (Table 5). For each amino acid, the coefficient of determination (r2) was >0.99 where the back calculated concentrations of the calibrators were within ±15% of the nominal value, except for the LLOQ (within ±20%). At least 75% of the calibrators, fulfilled this criterion.
Table 5. Linearity of the Kairos Amino Acid Kit in solution.
Sensitivity was assessed by the analysis of samples, prepared in 0.1 M HCl over the range 1–20 µM, in replicates of five on one day. The LOQ for each analyte was determined when the signal to noise ratio of the analyte peak was ≥10:1 and the precision performance of the replicates was ≤20%CV. The LOQ for all 20 amino acids analysed were ≤4 µM and are shown in Table 6.
Table 6. Sensitivity of the Kairos Amino Acid Kit in solution.
Accuracy was assessed through the analysisof the independently prepared panel samples as well as NIST SRM 2389a which was gravimetrically diluted to a concentration of 400 µM prior to analysis. All panel samples were ≤ ±12.3% bias (range -10.2 to +12.3%) and the diluted NIST SRM 2389a calculated concentrations were ≤ ±8.9% bias (range -8.9 to +6.2%) as shown in Table 7.
Table 7. Accuracy of the Kairos Amino Acid Kit in solution.
Conclusion
Kairos Amino Acid Kit enables biomedical researchers to achieve trusted results through a single flexible kit that allows them to accurately quantitate 20 amino acids within their normal physiological range in less than10 minutes using the ACQUITY UPLC/ACQUITY QDa System.
Using AccQ•Tag Ultra “3X” Derivitization Kit to derivatize the samples allows for a simple, robust, and fast reverse phase UPLC analysis of the amino acids without the need for mobile phase buffers or ion-pair reagents. This means that you do not need a dedicated system and sothe ACQUITY UPLC/ACQUITY QDa System can be used for other analyses. The derivatized samples are stable and in less than 10 minutes, the chromatographic conditions deliver the separation of isobaric amino acids giving you confidence in peak identification.
The method described in the research only Kairos Amino Acid Kitallows for the analysis of 20 amino acids in solution using anACQUITY UPLC/ACQUITY QDa System with good precision,linearity, sensitivity, and accuracy.